The paperless recorder enables recording of the test area temperature, humidity, and air exchange rate on a memory card (compact flash drive).
Model | VOC-1000 | VOC-2000 | ||
---|---|---|---|---|
Compact chamber | Temperature control range |
+20 to +100°C : At test operation (no load, with specimen) |
||
+80 to +250°C : Heat processing operation time (no load, no specimen) |
||||
Temperature and humidity control range | 30 to 90%RH (at+20 to +50°C)(no load, with specimen) | |||
Temperature and humidity control width | ±0.5°C / ±5% (no load, no specimen) | |||
Temperature and humidity distribution width*1 | ±1.0°C / ±5% (no load, no specimen) | |||
Background concentration | Less than 2 μg/m3 for single substance concentration | |||
Less than 20 μg/m3 for TVOC concentration | ||||
Air exchange rate | 0.2 to 2.0 times per hour (at test operation) | 0.2 to 1.5 times per hour (at test operation) | ||
Air exchange rate control width | ±5% (air exchange rate of 0.5 times per hour or higher) | |||
Mass transfer coefficient | 9 to 18m/h (water vapor conversion) | |||
Air tightness | Air leakage under 1% of supply air amount (at air exchange rate of 0.5 times per hour) | |||
Recovery rate | 80% or higher (Toluene) | |||
Effective inner dimension (W × H × D mm) |
1000 × 1100 × 810 | 2000 × 1100 × 810 | ||
(Inner capacity: Approx. 1 m3) | (Inner capacity: Approx. 2 m3) | |||
Material (Configuration) | Stainless steel plate (full circumference welded structure) | |||
Door effective dimension (W × H mm) | 1000 × 1100 | 2000 × 1100 | ||
Door lock | Latch lock (6 locations) | Latch lock (16 locations) |
Outside dimensions (W × H × D mm) | 2250 × 1950 × 1520 | 3250 × 1950× 1520 | ||
Door effective dimension (W × H mm) | 1600 × 1570 | 2600 × 1570 | ||
Air circulator | Cross line flow fan | |||
Refrigerator | Indoor air-cooled fully sealed (Refrigerant: R134a) | |||
Heater | Nichrome strip wire heater | |||
Evaporator | Multistage plate fin cooler | |||
Chamber weight | Approx. 1000 kg | Approx. 1800 kg | ||
Accessories | Recorder (paperless recorder), Integrating hour meter, Caster, Sampling port for specimen collection (2 locations), Air-circulating fan (VOC-020) | |||
Utility requirements | Power source facilities | Power supply |
AC200V 3φ 3W 50/60Hz Allowable voltage fluctuation : Rated voltage ±5% (The allowable operating range is the rated voltage ±10%) |
|
Maximum load capacity | 11.5kVA (15.8kVA when equipped with air compressor) | |||
Maximum current | 32.0A (45.6A when equipped with air compressor) | |||
Leakage breaker capacity | 40A (50A when equipped with air compressor) | |||
Compressed air | Discharged air flow | 160 L/min or highe (atmospheric pressure) Oil-free | ||
Control pressure | 0.6 to 0.9MPa | |||
Connection tube outer diameter | φ10mm | |||
Exhaust | Exhaust duct diameter | φ100mm (Flange diameter φ150mm) | ||
Discharge air volume | Max. 120 m3/h or higher | |||
Construction requirements |
|
|||
Supply water | Quality | Pure water (Quality equivalent to A1 water specified in JIS K0557) | ||
Maximum water supply rate | 20L (manual supply) | |||
Natural drainage pipe | Indirect drain VP30 1 units | |||
Floor load capacity | 350kg/m2 | 350kg/m2 |
*1 Measurement method complies with Japan Testing Machinery Association (JTM). However, the calculation method is half the maximum value minus the minimum value when stable.